Comparison of Standards of Olfactometry
Presented to EE-6 Odor Committee
at AWMA 1999 Annual Meeting, St. Louis, MO
prEN 13725 |
ASTM E-679-91 |
A&WMA EE-6 |
|
Requirements |
|||
Dynamic Dilution |
|||
Triangular Forced-Choice |
|
||
Ascending Concentration Series |
|||
Odor Free Dilution Air |
|||
Presentation Flow Rate (lpm) |
20 |
- - - |
8 |
Presentation Face Velocity (m/s) |
0.2 - 0.5 |
- - - |
0.02 - 0.05 |
Presentation Mask Diameter (cm) |
3 - 5 |
- - - |
6 - 10 |
Maximum Lower Dilution Ratio |
2^7 |
- - - |
2^3 |
Minimum Upper Dilution Ratio |
2^14 |
- - - |
2^13 |
Dilution Ratio Range |
2^13 |
- - - |
- - - |
Dilution Factor Between Levels |
1.4 - 2.4 |
2.0 or 3.0 |
2 |
Includes Sampling Requirements |
|||
Includes Olfactometer Performance Criteria |
|||
Includes Assessor Performance Criteria |
This standard will
supersede NVN 2820 (Netherlands), VDI 3881 (Germany) and all other European Standards.
Australia & New Zealand have drafted an equivalent standard.
* The CEN Draft also allows Binary Forced Choice, Yes/No Procedures, and Forced
Choice Probability Methods.